IMAGE QUALITY

SPI Exam Ultrasound Artifacts Cheat Sheet: Identification & Fixes

September 15, 202612 min read

Answer capsule: Ultrasound artifacts are image findings that do not accurately represent anatomy because an imaging-system assumption has been violated. For the SPI exam, identify the visual pattern, name the physical cause, and know the adjustment that can reduce it.

Ultrasound artifacts are a frequent SPI exam topic. Connect the visual pattern to a violated assumption: sound travels in a straight line, follows a direct path, travels at 1,540 m/s in soft tissue, and returns echoes from the expected location.

Reverberation versus comet-tail versus ring-down

Reverberation occurs when sound bounces repeatedly between strong reflectors. It commonly appears as multiple, equally spaced parallel echoes that become weaker with depth.

Comet-tail artifact is a short, bright, tapering reverberation-type artifact behind a small, highly reflective interface.

Ring-down artifact is different. It is a continuous echogenic band associated with resonance involving gas bubbles rather than ordinary discrete reverberation lines.

Suggested alt text: “Comparison of equally spaced reverberation echoes and a short tapering comet-tail artifact behind a bright reflector.”

Common SPI ultrasound artifacts

Artifact identification and cause reference
ArtifactAppearanceCause
ReverberationRepeated equally spaced linesMultiple reflections between strong reflectors
Comet-tailBright, short, tapering trailClosely spaced reverberation behind a small reflector
Ring-downContinuous bright bandResonance involving gas bubbles
ShadowingDark region behind a structureStrong attenuation or reflection
EnhancementIncreased brightness behind a structureLow attenuation through fluid
Mirror imageDuplicated anatomy across a strong reflectorAn indirect reflected path is interpreted as straight
RefractionDisplaced or duplicated anatomyBeam bending at an oblique interface
Side-lobe or beam-widthFalse echoes inside anechoic structuresEnergy from secondary or wider beam paths

Doppler aliasing as an artifact

Aliasing occurs in pulsed Doppler when the Doppler frequency shift exceeds the Nyquist limit, which is one-half of the pulse repetition frequency. The display may show flow in the wrong direction or wrap around the velocity scale.

Use the Nyquist limit calculator to connect imaging depth with PRF and the Nyquist limit. Also review the SPI physics formula sheet and ultrasound glossary.

Frequently asked questions

What is the difference between reverberation and comet-tail artifact?

Reverberation usually appears as repeated echoes between strong reflectors, often with equally spaced lines. Comet-tail artifact is a shorter, tapering reverberation-type artifact behind a small, highly reflective interface.

What is ring-down artifact?

Ring-down is a continuous echogenic band commonly associated with resonance involving gas bubbles. It is distinct from ordinary discrete reverberation lines.

What causes posterior acoustic shadowing?

Posterior acoustic shadowing occurs when a structure strongly attenuates or reflects the ultrasound beam, leaving reduced echoes behind the structure.

Ready to put this into practice? SonoPrep's exam simulator and flashcard system are built around the exact domains covered in this guide.