ARDMS SPI Study Reference

ARDMS SPI Ultrasound Physics Formula Sheet

The ultimate equation guide for the ARDMS Sonography Principles & Instrumentation examination. Need to test calculations? Check our interactive physics calculators.

Pulsed Wave & Resolution

ParameterFormulaKey Clinical Concept
Axial ResolutionSPL / 2LARRD. Determined by spatial pulse length. Lower value = better resolution.
Spatial Pulse Length (SPL)Number of Cycles × Wavelength (λ)Determined by both sound source and medium.
Pulse Duration (PD)Number of Cycles × Period (T)Time from start of a pulse to end of that pulse.
Duty Factor (DF)(Pulse Duration / PRP) × 100Typical clinical imaging range is 0.1% to 1.0%.

Doppler & Hemodynamics

ParameterFormulaKey Clinical Concept
Doppler EquationFd = (2 × Fo × v × cos θ) / cMaximum Doppler shift occurs at 0° when cos θ equals 1.0.
Nyquist LimitPRF / 2Threshold above which spectral aliasing occurs.
Poiseuille's LawΔP = (8 × η × L × Q) / (π × r⁴)A change in radius has an exponential effect on flow resistance.

Time, Distance & Frame Rate

ParameterFormulaKey Clinical Concept
Period (T)1 / Frequency (f)Period and frequency are reciprocals. Set by the sound source.
Wavelength (λ)Propagation Speed (c) / Frequency (f)In soft tissue (1540 m/s), λ in mm = 1.54 / frequency in MHz.
Pulse Repetition Period (PRP)1 / PRFTime from the start of one pulse to the start of the next, including listening time.
Go-Return Time (Time of Flight)13 µs per cm of depthRule of thumb for soft tissue: an echo from 1 cm depth returns in about 13 µs.
Maximum PRF for a Depth77,000 / depth (cm) (Hz)Equivalent to 77 / depth (cm) in kHz. Deeper imaging forces a lower PRF.
Frame RatePRF / Lines per FrameMore lines, deeper depth or multiple focal zones lower the frame rate.

Decibels, Intensity & Reflection

ParameterFormulaKey Clinical Concept
IntensityPower / AreaDoubling the power doubles the intensity; halving the beam area also doubles it.
Decibels (intensity or power)dB = 10 × log₁₀(I₂ / I₁)+3 dB ≈ ×2, −3 dB ≈ ×½, +10 dB = ×10. For amplitude or pressure the multiplier is 20 × log₁₀.
Total AttenuationAttenuation Coefficient (dB/cm) × Path Length (cm)Path length is the round-trip distance for an echo (2 × depth).
Half-Value Layer (HVL)3 dB / Attenuation Coefficient (dB/cm)Depth at which intensity has fallen by 3 dB. Shallower for higher frequencies.
Intensity Reflection Coefficient (IRC)[(Z₂ − Z₁) / (Z₂ + Z₁)]²Normal incidence. Larger impedance mismatch means more reflection; equal impedances mean none.

Beam Geometry & Hemodynamics

ParameterFormulaKey Clinical Concept
Near Zone Length (Focal Depth)D² / (4 × λ)D is the aperture (crystal) diameter. Larger aperture or higher frequency lengthens the near zone.
Continuity EquationA₁ × v₁ = A₂ × v₂Velocity rises at a stenosis because the cross-sectional area falls.
Simplified BernoulliΔP = 4 × v²Pressure gradient in mmHg with velocity in m/s.
Reynolds Number(ρ × v × D) / ηValues above roughly 2000 predict turbulent flow; below that flow tends to be laminar.
Resistive Index (RI)(PSV − EDV) / PSVUses peak systolic and end-diastolic velocities.
Pulsatility Index (PI)(PSV − EDV) / Mean VelocityUses the mean velocity over the cardiac cycle.

Wave Properties & Attenuation

ParameterFormulaKey Clinical Concept
Acoustic Impedance (Z)Density (ρ) × Propagation Speed (c)Measured in Rayls. Determines boundary reflection magnitude.
Attenuation CoefficientFrequency (MHz) / 2In soft tissue, approximately 0.5 dB/cm/MHz.
Snell's Lawsin(θt) / sin(θi) = c₂ / c₁Refraction requires oblique incidence and different propagation speeds.